• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRLNE BLU 30IN 6-0 D/A C-1 EP; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRLNE BLU 30IN 6-0 D/A C-1 EP; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number EP8706H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a cabg procedure on (b)(6) 2024 and suture was used.During the procedure, the suture broke.Another like device was used.The procedure was completed with no adverse patient consequences.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: if possible, please confirm the number of sutures that broke during this procedure (how many from 6/0 prolene and how many from 7/0 prolene).5 eaches of 6/0 and 4 eaches of 7/0.Note: related events reported via 2210968-2024-04573, 2210968-2024-04574, 2210968-2024-04575, 2210968-2024-04576, 2210968-2024-04577, 2210968-2024-04578, 2210968-2024-04579, 2210968-2024-04580,2210968-2024-04581.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h4, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Date sent to the fda: 8/6/2024.A device has been received; however clarification is requested whether the product belongs to this complaint.The following information was requested: product code lot# uabeal and lot# tpbcmb/code ep8706.1.Could you please clarify if the additional device belongs to this complaint? 1a.If yes, did a device malfunction occur during the same procedure? 2.If not, could you please clarify if the additional received product belongs to a different complaint? if so, can you please provide the complaint number.3.If the device was not reported before, please provide more details of device malfunction.4.If the product doesn¿t belong to any complaint, please let us know so we can discard it or advise if the product is required to be returned.5.If you have the correct complaint product in your possession, please forward the product to ethicon as soon as possible.If you have already sent the product to ethicon, please reply to ethwcqcom@its.Jnj.Com with the ups, dhl or fedex tracker number.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 8/6/2024.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Three unopened samples were received for analysis.Product code ep8706h.This product code is double-armed.In order to evaluate the condition of the returned samples, the packets were opened.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand and no issues related to breakage sutures or anomalies were observed during the evaluation.The functional test was performed using instron equipment and the tensile force was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 30IN 6-0 D/A C-1 EP
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key19142540
Report Number2210968-2024-04573
Device Sequence Number7303619
Product Code GAW
UDI-Device Identifier10705031218949
UDI-Public(01)10705031218949(11)240119(17)281231(10)UABEAL
Combination Product (Y/N)N
Initial Reporter CountryHR
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEP8706H
Device Lot NumberUABEAL
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer04/23/2024
07/18/2024
07/18/2024
Initial Report FDA Received Date04/19/2024
Supplement Report FDA Received Date05/15/2024
08/07/2024
08/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-