• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL LEFT SIZE GH; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL LEFT SIZE GH; INSTRUMENT, KNEE Back to Search Results
Catalog Number 42517000707
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).D10: medical product: tibial articular surface provisional left size cd: catalog#42517000303, lot#64159154 the product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Upon completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a tibial articular surface provisional instrument fractured.There was no patient impact and no adverse events have been reported as a result of the malfunction.It was reported that no further information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h4; h6; h9; h10.H6 : suggested component code: tasp bottom- mechanical (g04) - provisional bottom.Visual examination of the provided picture identified both devices to be fractured.Fragments are shown in the picture.No further evaluation can be made.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.Investigation into this issue determined that the likely root cause for the tasp fractures is bending/torsional loading on the device.Ps tasp top components and standard (non-cps) tasp bottom components have been modified to prevent fracture.The fractured tasp component in this complaint was manufactured before the design modification.The root cause is considered to be a previously addressed design issue.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Upon review of our reporting decision on fracture of articular surface provisionals.A review of all complaints for this product indicates that there has never been an event that has resulted in a death or serious injury.Therefore, zimmer biomet will not be reporting this failure mode going forward.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA TIBIAL ARTICULAR SURFACE PROVISIONAL LEFT SIZE GH
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19143227
Report Number0001822565-2024-01335
Device Sequence Number18108206
Product Code JWH
UDI-Device Identifier00889024238459
UDI-Public(01)00889024238459(10)64159154
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number42517000707
Device Lot Number62601574
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer07/03/2024
Initial Report FDA Received Date04/19/2024
Supplement Report FDA Received Date07/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2578-2014
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Patient SexPrefer Not To Disclose
-
-