| Model Number |
978B128 |
| Medical Device Problem Codes |
Failure to Deliver Energy (1211); Migration or Expulsion of Device (1395); Unintended Collision (1429); Energy Output Problem (1431); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/27/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for urinary dysfunction (incontinence, urgency, and/or frequency) it was reported that patient fell in january and then couldn't increase stim past 0.7 and their symptoms got worse from the loss of therapy.Rep interrogated device and found the same issues, impedance at electrodes 1 and 3.Imaging showed lead was also dislodged so hcp replaced with new lead.
|
| |
|