| Model Number |
8888135131 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, the patient with chronic kidney disease (stage 5) went to the catheterization room for right internal jugular vein catheterization at 13:30.The surgery went smoothly, the depth of vein insertion was 13.5 cm (centimeter), and the exposed length of the catheter was 0 cm.There was no bleeding or exudation at the puncture site, and there was no hematoma on the surrounding skin.The dressing was dry and clean, and ice was applied to the place where the catheter was inserted.The patient was escorted to the radiology department for a chest x-ray examination and then returned to the ward.Hemodialysis treatment was performed at 15:00.1 day after the catheter had been in place, during the evaluation before getting on the hemodialysis machine, it was found that there was a little blood stain on the pipeline which meant extension tube.The exact location of the leak on the catheter was the blue extension tube near the adapter.No new blood stain was found after disinfection.After the hemodialysis machine was put on, the pipeline was wrapped with 75% alcohol gauze.When the hemodialysis machine was put off at 17:38, a small amount of blood was found on the dressing.The inspection revealed that there was a small hole in the pipeline and leaking blood.The patient was undergoing hemodialysis, and the risk of blood infection would increase if there was a small hole of blood leakage in the dialysis pipeline.Nothing unusual observed on the device prior to use.No cleaning agent used on the device.The clamp was not moved periodically.Flushing was done prior to use and no exception found.The product/tube was replaced with the same product id (identifier) on the same day as remedial action performed and the treatment was completed.A few milliliters the amount of blood lost.Blood transfusion was not required.There was no reported patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the patient with chronic kidney disease (stage 5) went to the catheterization room for right internal jugular vein catheterization at 13:30.The surgery went smoothly, the depth of vein insertion was 13.5 cm (centimeter), and the exposed length of the catheter was 0 cm.There was no bleeding or exudation at the puncture site, and there was no hematoma on the surrounding skin.The dressing was dry and clean, and ice was applied to the place where the catheter was inserted.The patient was escorted to the radiology department for a chest x-ray examination and then returned to the ward.Hemodialysis treatment was performed at 15:00.1 day after the catheter had been in place, during the evaluation before getting on the hemodialysis machine, it was found that there was a little blood stain on the pipeline which meant extension tube.The exact location of the leak on the catheter was the blue extension tube near the adapter.No new blood stain was found after disinfection.After the hemodialysis machine was put on, the pipeline was wrapped with 75% alcohol gauze.When the hemodialysis machine was put off at 17:38, a small amount of blood was found on the dressing.The inspection revealed that there was a small hole in the pipeline and leaking blood.The patient was undergoing hemodialysis, and the risk of blood infection would increase if there was a small hole of blood leakage in the dialysis pipeline.Nothing unusual observed on the device prior to use.No cleaning agent used on the device.The clamp was not moved periodically.Flushing was done prior to use and no exception found.The product/tube was replaced with the same product id (identifier) on the same day as remedial action performed and the treatment was completed.A few milliliters the amount of blood lost.Blood transfusion was not required.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted that there was a cut in the extension tube which led to a leak.It was reported that there was a leak on the extension tube near the luer adapter.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur by clamping the tubing too close to the luer adapter.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: clamping repeatedly on the same spot could weaken the tubing and clamping near the adapter and hub should be avoided.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|