• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC*SIGMA C/R NPOR FE RT SZ2.5; SIGMA KNEE PRIMARY : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC*SIGMA C/R NPOR FE RT SZ2.5; SIGMA KNEE PRIMARY : KNEE FEMORAL Back to Search Results
Catalog Number 960018
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Article entitled ¿3d polyethylene wear analysis after primary cruciate-retaining tka failure using optical scanning and cad models references¿ written by matej valic, vesna leva¿ic, ingrid milo¿ev, mateja blas, eva podov¿ovnik, jaka koren, and rihard treb¿e was reviewed.Our purpose was to identify (a) whether volumetric wear (adjusted for time in-situ) significantly differ among groups of common causes of tka failure, (b) whether there is a characteristic wear distribution pattern for a common cause of tka failure, and (c) whether nominal insert size (small vs large) and components size ratio (femur-to-insert) influence on volumetric wear rates.The study included patients that had been implanted with a sigma tka (cemented tibial tray and porous coated femoral component).Within the patient that met the criteria, there were noted to be 86 revision cases due to failure (no mention of why).59/86 tibial inserts were retrieved.2 of the 59 inserts exhibited inexplicable abnormal type of damage and were excluded from the final analysis.Final cohort included 57 inserts.There was no mention of cement mfg or patella implants.Results: 86 revisions.Only 5 patients had a single detected causes, while the remaining had 2-3 detected causes.35 cases of infection (13 of the 35 cases were noted to have loosening as well), 18 cases of osteolysis and loosening (unknown device), 6 cases of osteolysis, 28 cases of joint malignment or component malpositioning, 12 cases of instability, 10 cases with ligament insufficiency, and 3 cases for other isolated causes.Volumetric and linear wear noted on the 57 liners.2 liners excluded from analysis are pictured in figure 5 ¿ show signs of wear.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SIGMA KNEE PRIMARY : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19144899
Report Number1818910-2024-08679
Device Sequence Number7444223
Product Code JWH
UDI-Device Identifier10603295232070
UDI-Public10603295232070
Combination Product (Y/N)N
Initial Reporter CountrySI
PMA/510(K) Number
K952830
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Literature,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number960018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer04/22/2024
Initial Report FDA Received Date04/19/2024
Supplement Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-