| Catalog Number |
MS18597 |
| Medical Device Problem Code |
Inaudible or Unclear Audible Prompt/Feedback (2283)
|
| Health Effect - Clinical Code |
Asystole (4442)
|
| Date of Event |
03/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that there was no alarm heard for an asystole on a delta device.No additional information regarding the event was provided.There was no report of patient injury or death.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation has just started; results will be provided in a follow-up report.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported to draeger that the asystole alarms were not heard on the delta monitor.The draeger engineer that serviced the delta monitor at the customer's site confirmed the cause of the reported behavior was the result of the or locked option being enabled by the user, which causes the alarms to be silenced.Testing of the delta monitor did not indicate any malfunctions as the device was found to be operating as intended.The cited delta monitor has been returned to use for patient monitoring.Root cause was identified as user error.No further issues have been reported.
|
| |
|
Event or Problem Description
|
|
The customer reported that there was no alarm heard for an asystole on a delta device.No additional information regarding the event was provided.There was no report of patient injury or death.
|
| |
|