| Model Number |
DEX900S13 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Dysphagia/ Odynophagia (1815); Dyspnea (1816); Extreme Exhaustion (1843); Headache (1880); Nausea (1970)
|
| Date of Event |
04/10/2024
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging exhaustion, nausea, headaches, shortness of breath, choking attacks, difficulty swallowing and death.In addition, patient alleges filters were completely black every morning.There was no report of medical intervention.No additional information can be requested at this time.The manufacturer was made aware of this complaint through a representative of the customer.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 other text : device not returned to manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: evaluation code grid: from device not returned to type of investigation not yet determined.Evaluation results code grid: from no findings available to results pending completion of investigation conclusion code grid: from cause not established to conclusion not yet available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to a third-party service center on (b)(6) 2024, but the investigation is still ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
As per device evaluation received on 09/03/2024: the device "dreamstation autosv" was returned to the third party service center for evaluation.During the evaluation of the device at the third-party service center, the device was visually inspected and foam particles were not observed inside the assembly.In addition, dust/dirt was found inside the device.Additionally, the patient¿s complaint was not confirmed, and the device was scrapped.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.In this report, box d, h has been updated/corrected.
|
| |
|