Catalog Number 19AGN-751
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2024, a 19mm sjm regent heart valve was chosen for a aortic valve replacement (avr) procedure.During preparation, there was nothing wrong with the valve.The valve was implanted in the patient.After the procedure, once the valve was implanted it was found that the valve leaflet couldn't move.The physician removed the valve and a new 19mm non-abbott valve was implanted intra-procedurally, while the patient was still on bypass.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient was reported stable.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
An event of physical resistance/sticking of the leaflets was reported and did not confirmed.Morphological and hydrodynamic examination indicated the valve met abbott specifications.Hydrodynamic testing upon return to abbott and at the time of manufacturing indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key19146344
Report Number2135147-2024-01778
Device Sequence Number7443783
Product Code LWQ
UDI-Device Identifier05414734005869
UDI-Public(01)05414734005869(17)251105(10)7731655
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/05/2025
Device Catalogue Number19AGN-751
Device Lot Number7731655
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer06/26/2024
Initial Report FDA Received Date04/19/2024
Supplement Report FDA Received Date07/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age40 YR
Patient SexFemale
-
-