| Catalog Number |
19AGN-751 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a 19mm sjm regent heart valve was chosen for a aortic valve replacement (avr) procedure.During preparation, there was nothing wrong with the valve.The valve was implanted in the patient.After the procedure, once the valve was implanted it was found that the valve leaflet couldn't move.The physician removed the valve and a new 19mm non-abbott valve was implanted intra-procedurally, while the patient was still on bypass.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient was reported stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of physical resistance/sticking of the leaflets was reported and did not confirmed.Morphological and hydrodynamic examination indicated the valve met abbott specifications.Hydrodynamic testing upon return to abbott and at the time of manufacturing indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|