• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT ORTHOPEDICS INC. ADVANCE® II COCR TIBIAL BASE NONPOROUS SZ 3 STANDA; KNEE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT ORTHOPEDICS INC. ADVANCE® II COCR TIBIAL BASE NONPOROUS SZ 3 STANDA; KNEE COMPONENT Back to Search Results
Model Number KTCCNP30
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/13/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Allegedly, the patient was diagnosed with radiology.Evident loosening of the prosthesis was detected, the tibial base was replaced.The initial implantation was performed in 2015.First revision: it was detected shortly after implantation that there was infection and an insert change was performed.Pe was not firmly attached to the tibial component as intended, but became detached from it.As a result, there was wear of the underside of the insert with polyethylene granulomas.There were polyethylene granulomas throughout the joint and complete loosening of the prosthesis.The joint was reconstructed with a complex coupled, modular, stem-guided prosthesis and all granulomas were removed in approximately 3 hours of revision surgery.The explants were given to the patient: product id: kimp312r advance® ii medial pivot tibial insert sz 3 right 12mm / lot no.: 1687317/ qty: (b)(4).Product id: kftcnp3r advance® primary femoral size 3 right nonporous / lot no.: 1707334/ qty: (b)(4).Product id: ktccnp30 advance® ii cocr tibial base nonporous sz 3 standard/ lot no.: 1644600/ qty:(b)(4).
 
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE® II COCR TIBIAL BASE NONPOROUS SZ 3 STANDA
Common Device Name
KNEE COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
rachael wise
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key19146657
Report Number3010536692-2024-00266
Device Sequence Number7920856
Product Code HRY
UDI-Device IdentifierM684KTCCNP301
UDI-PublicM684KTCCNP301
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K960617
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberKTCCNP30
Device Catalogue NumberKTCCNP30
Device Lot Number1644600
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/19/2024
Supplement Report FDA Received Date10/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexMale
-
-