| Model Number |
IBAL TKA,TIB BEARING TRL SZ 8, 12MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, it was reported by a sales representative via sems (b)(4) that an ar-623-t812 ibalance tka, tibial bearing trial size 8 was broken during use with no further information provided.There was no effects to the patient reported.This was discovered during a procedure.Additional information was received on 3/29/2024: the impactor was broken when the tibial implant was implanted; the poly trial broke when attempting to trial the sizes.The device broke while being put into the patient, and all fragments were retrieved from the patient.The case was delayed by about a minute when switching to another poly trial size.The case was completed with another trial and finished normally with no adverse effects on the patient.This was discovered during a total knee arthroplasty procedure on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 the complaint allegation is confirmed.Based on the image provided from the field, it is evident that the trial has damage around the edges and a piece broken off in the center.Consequently, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to user-applied mechanical forces during insertion.
|
| |
|
Search Alerts/Recalls
|