ETHICON INC. EB EXC GRN 4X30IN 5 S/A V-37; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
|
Back to Search Results |
|
| Catalog Number |
MB66G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 device not returned additional information was requested, and the following was obtained: * does the needle piece retain in the patient's tissue? no * were x-rays performed to localize the needle piece? n/a * if retained, were there any patient consequences? please specify.N/a * was the needle piece removed or is it planned to be removed? if yes, please provide details.* please clarify how many devices will be returned for analysis? 1 needle.Other needle was discarded.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Related events captured via: 2210968-2024-04626 2210968-2024-04627.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an orthopedic procedure on an unknown date; and a suture was used.During the procedure, the suture broke during use.Another like device was used to complete the procedure.There were no adverse consequences to the patient.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).A review of the batch manufacturing records was conducted, and no related non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Additional information: d9.H3 evaluation: the product received for analysis was identified as product code.Visual inspection and metallurgical analysis were performed on the returned product.A fracture was observed at the suture attachment of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.The needle was noted with marks that appears to be by surgical instrument.A scanning electron microscope (sem) was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation of the sample revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage observed coincidental to the fracture provide additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the attachment area of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.The needle breakage was observed.A review of the batch manufacturing records was conducted, and no related non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|