-patient medications: pepcid, amlodipine, plavix, valium, zetia, synthroid, lisinopril, toprol, minoxidil, and repatha.The gore® excluder® aaa endoprosthesis instructions for use (ifu) states that adverse events that may occur and/or require intervention include but are not limited to endoleak.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
The field sales associate (fsa) reported the following: on (b)(6) 2024, the patient was implanted with a gore® excluder® aaa endoprostheses to treat an abdominal aortic aneurysm.The patient tolerated the procedure.It was reported that on (b)(6) 2024, the patient had a reintervention to treat a proximal type i endoleak.Reportedly the patient has a calcified aorta, and this was causing the endoleak.The physician used thirteen endoanchors, the endoleak was resolved.There was no aneurysm sac enlargement.The patient tolerated the reintervention.
|