AIZU OLYMPUS CO., LTD. EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-N180
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during reprocessing, the gastrointestinal videoscope tested positive for 7 colony forming units (cfus)/100ml of aerobic microorganisms.A subsequent test was performed and tested positive for 8 cfu's/100ml of aerobic microorganisms.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Additional Manufacturer Narrative
The device evaluation is ongoing.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacture's (lm) review of the customer completed cleaning disinfection and sterilization (cds) checklist, results of third-party testing, the device evaluation and the lms final investigation.The lm reviewed the customer provided the cds processes where no obvious deviations from instructions for use (ifu) were identified.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: on (b)(6) 2024 sampling from: all channels cfu: 2 cfu bacterial identification: bacillaceae the device was evaluated where no abnormalities were found that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the relevance between the device and the detection of bacteria could not be established and a root cause could not be determined.The following is included in the device ifu: ¿an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19149342
Report Number9610595-2024-08326
Device Sequence Number12623393
Product Code FDS
UDI-Device Identifier04953170307508
UDI-Public04953170307508
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGIF-N180
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer05/27/2024
Initial Report FDA Received Date04/19/2024
Supplement Report FDA Received Date06/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-