SMITH & NEPHEW, INC. REDAPT SLVD MONO STEM 240MM SZ 22 HO; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 71354829
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: case-2024-00205223-1.This complaint was opened by smith+nephew to document a patient complication identified through the review of clinical evidence from post market clinical data collection activities that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, during a right tha conversion performed on an unknown date due to failed right hip arthroplasty with history of infection, severe bone loss and abductor deficiency, the patient sustained an intraoperative femoral fracture at the tip of the stem.No further information is available.
 
Additional Manufacturer Narrative
H3, h6: although the involvement of all the devices was reported, the relationship with the investigated failure was directly associated to the redapt stem.Given the nature of the alleged incident, the device could not be returned for evaluation; therefore, a device analysis could not be performed.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for total hip systems revealed in the warnings and precautions section that congenital deformity, improper implant selection, improper broaching or reaming, osteoporosis, bone defects due to misdirected reaming, trauma, strenuous activity, improper implant alignment or placement, patient non compliance, etc.Can increase risk of femoral or pelvic fractures.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include patient bone quality and/or surgical technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Event or Problem Description
It was reported that, during a right tha conversion performed on an unknown date due to failed right hip arthroplasty with history of infection, severe bone loss and abductor deficiency, the patient sustained an intraoperative comminuted femoral fracture at the tip of the stem with lateral exposure.The proximal body was prepared and the stem was placed.After this, the fracture was reduced around the stem and cabled.No further information is available.
 
Additional Manufacturer Narrative
Corrected data: b5 (narrative), h6 (health effect - impact code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDAPT SLVD MONO STEM 240MM SZ 22 HO
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19149506
Report Number1020279-2024-00793
Device Sequence Number15340530
Product Code LZO
UDI-Device Identifier00885556580165
UDI-Public00885556580165
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K211176
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Study,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71354829
Device Lot Number16KTM0063E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer06/14/2024
07/09/2024
Initial Report FDA Received Date04/19/2024
Supplement Report FDA Received Date06/17/2024
07/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/05/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PN: 71342800 / LN: 18GM01046; PN: 71354223 / LN: 13FM08183A; PN: 71354238 / LN: 17FM15724; PN: 75018960 / LN: B1510783
Outcome Attributed to Adverse Event Other;
Patient Age41 YR
Patient SexMale
-
-