| Model Number |
25MM CENTRAL SCREW, MODULAR |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 3/26/2024, it was reported by a sales representative via email that an ar-9560-24-2 modulas baseplate disassociated from an ar-9561-25s central screw during insertion.This was discovered during a rtsa procedure on (b)(6) 2024.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to improper/shallow engagement of the mating devices.
|
| |
|
Search Alerts/Recalls
|