| Model Number |
TFAT30 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Prolapse (2475)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A non-health care professional reported that following an intraocular lens (iol) implant procedure, the patient experienced decline in vision of right eye.The lens was exchanged for another lens model 2 years 7 months following the initial procedure.Clinical reason for explant was mentioned as mechanical complication with vitreous prolapse of right eye.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3.,h.6 and h.11.The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint.Each intra ocular lens (iol) is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the company model (1.50cyl), 24.0 diopter (add power +2.2/+3.2) lens was replaced with an unspecified atiol(advanced technology intra ocular lens) model due to a reported decline in vision and vitreous prolapse.The cause of the vitreous prolapse was not provided.The product has not been received to evaluate.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to requests for follow-up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|