| Model Number |
15 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.Stryker evaluated the customer¿s device but was unable to duplicate the reported issue.After observing proper device operation through functional and performance testing the device was returned to the customer for use.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device was not reading an ecg signal through the paddles lead and displayed dashed lines.In this state defibrillation therapy may be delayed if needed or may lead to use error resulting in a failure to deliver defibrillation.A backup device was available and was used.This issue is patient related; however, there was no adverse patient outcome.
|
| |
|