| Catalog Number |
MS26800 |
| Medical Device Problem Codes |
No Audible Prompt/Feedback (2282); No Visual Prompts/Feedback (4021)
|
| Health Effect - Clinical Code |
Tachycardia (2095)
|
| Date of Event |
03/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the infinity central station failed to alarm for ventricular tachycardia on (b)(6) and (b)(6).There was no report of patient injury or patient death.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation has just started; results will be provided in a follow-up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Draeger engineering found there were no log entries for ventricular tachycardia (vt) for the date of the reported event.However, there were three svr arrhythmia alarms found in the ecg-strips that were generated on the date of event ((b)(6) 2024).Requests were made for the arrhythmia settings, additional logs, along with additional information but they were not provided for this investigation.Therefore, the specific root for this issue could not be determined.The software versions on ics and the m300 are old.Various enhancements have been made to the new software versions regarding alarm detection and should be considered for future monitoring.This complaint is being closed and if in the future the proper logging and information is provided, a new complaint will be opened for the new information.There have been no additional reports of any errors from this customer's site.
|
| |
|
Event or Problem Description
|
|
The customer reported that the infinity central station failed to alarm for ventricular tachycardia on march 5th and april 8th.There was no report of patient injury or patient death.
|
| |
|