US SURGICAL PUERTO RICO TA; STAPLER, SURGICAL Back to Search Results
Model Number TA9048L
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during an open procedure, the device partially fired.A new device was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant product: unknown ta, unknown ta (lot#unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: e1(first name, last name), e3, e4, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the single use loading unit (sulu) noted that all of the staples were partially advanced in the cartridge.The staples were in usable unformed condition.The instrument was not received.It was reported that the staple line was incomplete.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when the firing stroke is not completed during application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: to fire the instrument, squeeze the handle a second time until it reaches a solid stop and it locks in the back position.If the instrument handle is partially squeezed during firing then released, bleeding may occur as the instrument was not fully fired.The black release button may be used to open the instrument at any time during approximation.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TA
Common Device Name
STAPLER, SURGICAL
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19153192
Report Number2647580-2024-01753
Device Sequence Number8445499
Product Code GAG
UDI-Device Identifier10884521043268
UDI-Public10884521043268
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K231491
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberTA9048L
Device Catalogue NumberTA9048L
Device Lot NumberP1A1365
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2024
Supplement Date Received by Manufacturer06/05/2024
Initial Report FDA Received Date04/22/2024
Supplement Report FDA Received Date06/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE NOTE ON H11
Patient SexUnknown
-
-