• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL RBC / PLASMA, ISBT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL RBC / PLASMA, ISBT Back to Search Results
Catalog Number 4825002
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process, a follow-up report will be provided.
 
Event or Problem Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.Wbc count is not available at this time.The platelet collection set is not available for return because it was discarded by the customer.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in b.5, h.6 and h.11.Investigation: the customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.The lrs chamber has a fixed volume and therefore holds a fixed number of white blood cells for a given chamber flow.On infrequent occasions, this holding capacity may be exceeded, causing higher than expected white blood cell content within the product.This may result in a collected product with greater than expected wbc count.The trima machine may give a message to verify the wbc content in the product or may provide no message to the operator depending on the circumstances.Root cause: there are various reasons for elevated wbc levels in collected product such as donor factors, kit misloads, airblocks, disposable set deviations, paused centrifuge, or other processing difficulty in maintaining the steady state necessary to efficiently separate blood components.Regarding donor factors, there appears to be patterns with repeat donors that produce platelet products with high wbc counts.Saturation events leading to elevated wbcs in collected product are also possible if the donor pre-counts are entered falsely much lower than actual as this can cause the lrs chamber to fill before the system expects and wbcs to escape before detection.Misloading of a set or disposable set deviation may cause a plasma line occlusion or an air block within the centrifuge which may lead to the white blood cells getting flushed out of the lrs chamber into the product.
 
Event or Problem Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.Wbc count is not available from the customer the platelet collection set is not available for return because it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA ACCEL RBC / PLASMA, ISBT
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key19153701
Report Number1722028-2024-00149
Device Sequence Number8463972
Product Code GKT
UDI-Device Identifier05020583825000
UDI-Public05020583825000
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
BK190332
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2025
Device Catalogue Number4825002
Device Lot Number2307182142
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2024
Supplement Date Received by Manufacturer06/06/2024
Initial Report FDA Received Date04/22/2024
Supplement Report FDA Received Date06/07/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-