| Model Number |
0684-00-0479-07 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional event site information : (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that when the intra-aortic balloon (iab) was inserted, it would not inflate.A new iab was inserted to provide therapy.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint #(b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The iab was returned with the membrane completely unfolded.No blood was visible on the catheter.The extender tubing was also returned.An underwater leak test of the balloon, catheter, y-fitting, extracorporeal and extender tubing was performed and no leaks were detected.The iab was placed on the cardiosave pump and the iab fully inflated.No alarm sounded from the pump.The reported event cannot be confirmed by the evaluation.A lot history record review was completed for the reported product.No non-conformances were found that are considered to be related to the event.Reference complaint #(b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only** providing updated device identification information in alignment with gudid.Complaint record id (b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|