MAUDE Adverse Event Report: DATASCOPE CORP. LINEAR 7.5FR. 34CC IAB (INDIA); SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. LINEAR 7.5FR. 34CC IAB (INDIA); SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0479-07
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional event site information : (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event or Problem Description
It was reported that when the intra-aortic balloon (iab) was inserted, it would not inflate.A new iab was inserted to provide therapy.There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint #(b)(4).
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The iab was returned with the membrane completely unfolded.No blood was visible on the catheter.The extender tubing was also returned.An underwater leak test of the balloon, catheter, y-fitting, extracorporeal and extender tubing was performed and no leaks were detected.The iab was placed on the cardiosave pump and the iab fully inflated.No alarm sounded from the pump.The reported event cannot be confirmed by the evaluation.A lot history record review was completed for the reported product.No non-conformances were found that are considered to be related to the event.Reference complaint #(b)(4).
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
**udi related data quality updates only** providing updated device identification information in alignment with gudid.Complaint record id (b)(4).
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5FR. 34CC IAB (INDIA)
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE CORP.
15 law drive
fairfield NJ 07004
Manufacturer Contact
jingal amin
15 law drive
fairfield, NJ 07004
MDR Report Key19153972
Report Number2248146-2024-00242
Device Sequence Number7861916
Product Code DSP
UDI-Device Identifier10607567109695
UDI-Public(01)10607567109695(10)3000349897(17)261117(11)231117
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K041281
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0479-07
Device Catalogue Number0684-00-0479-07
Device Lot Number3000349897
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2024
Supplement Date Received by Manufacturer05/10/2024
06/07/2024
11/20/2024
Initial Report FDA Received Date04/22/2024
Supplement Report FDA Received Date05/10/2024
06/07/2024
11/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
Patient Weight70 KG
-
-