MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG POLARSTEM DETACHABLE RASP 01; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG POLARSTEM DETACHABLE RASP 01; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 75023019
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 03/25/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal reference number: case (b)(4).
 
Event or Problem Description
It was reported that during a thr surgery, the patient experienced a fracture of the femur due to the dullness of one (1) polarstem detachable rasp 01, which failed to properly broach the bone due to compacting.It is unknown how this adverse event was treated.The procedure was resumed after a significant delay, using the same device.The patient's current health status is unknown.
 
Additional Manufacturer Narrative
H3, h6: it was reported that during a total hip replacement surgery, the patient experienced a fracture of the femur due to the dullness of one (1) polarstem detachable rasp 3, which failed to properly broach the bone due to compacting.The device intended for use in treatment was not returned for investigation.A product evaluation was not possible.No batch number was communicated so the production history review was not possible.Due to insufficient information it is not possible to perform a review of past corrective actions.As the batch number is unknown, it is not possible to perform a complaint history review at batch level, and the complaint history review based on the product number revealed 6 additional similar complaints.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.Insufficient information was made available to determine the revision of the instructions for use applicable to the device at the time of manufacturing.However, a review of the current revision was conducted and it revealed that the instructions for use (lit.No.12.23, ed.03/21) states "hip fracture" as a ¿potential adverse device effect¿ in combination with the implantation of a hip prosthesis.No clinically sufficient data was provided to perform a medical investigation.The performed investigation does not lead to an accurately determined cause.Based on the available information it is not possible to investigate whether the reported device met manufacturing specifications upon release for distribution.Due to insufficient information, it is not possible to indicate factors which could have contributed to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.There is no need for further actions because of the limited information provided.Nevertheless, smith + nephew will continue to monitor this device for similar issues.This complaint will be reopened should additional information or the device be received.Additional information: d5, d7a, d8, g2, h8 corrected data: b2, b5, h6 (health effect - impact code).
 
Event or Problem Description
It was reported that during a thr surgery, the patient experienced a fracture of the femur due to the dullness of one (1) polarstem detachable rasp 3, which failed to properly broach the bone due to compacting.This adverse event was treated by using an accord cable.The procedure was resumed after a significant delay, using the same device.The patient's current health status is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARSTEM DETACHABLE RASP 01
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19154065
Report Number9613369-2024-00044
Device Sequence Number20702400
Product Code JDI
UDI-Device Identifier07611996097511
UDI-Public07611996097511
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number75023019
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2024
Supplement Date Received by Manufacturer05/07/2024
Initial Report FDA Received Date04/22/2024
Supplement Report FDA Received Date05/08/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-