| Model Number |
SMALL CAGED GLENOID |
| Medical Device Problem Codes |
Fracture (1260); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Loss of Range of Motion (2032); Metal Related Pathology (4530); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/27/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, the 72 y/o female patient had complained of pain for approximately 4 years (implants were put in 7-8 years ago).However, due to covid it had not been possible to x-ray/ct the patient so the surgeon is not fully clear what route of deterioration had been taking place.Patient first stage revision of a caged glenoid anatomic.All implants removed and replaced with a spacer.The patient's rotator cuff failed.Patient lost function.During the surgery the 3 x peripheral pegs and attached polyethylene were retrieved.The caged glenoid appeared to have significantly 'eroded' with possible erosion after being lodged under the humeral head.The medial calcar had also completely disappeared due to possible stress shielding.There was significant metalosis observed leading to the belief that this could have been the cage.There was no evidence of the cage being anywhere else (as per x-rays and ct's).The surgeon explored all reasonable options to find it but there was no sight of it.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Sales rep was unable to obtain photos and x-rays.The devices are not available for evaluation due to it is not hospital policy to release implants.Request has been made none the less.If approved, they will be collected and sent to gv.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d10: concomitant products 41mm head.4.5mm replicator plate.11mm humeral stem.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections: h6: health effect / clinical code.Additional information: d9 - device return.H3 - the revision reported was likely the result of failure of the patient¿s rotator cuff and 4 years of reported pain.The glenoid component was returned fractured with the center cage missing.The cause and sequence of events cannot be determined because no sequential, pre-revision radiographs were provided.H6 - medical device problem code, health effect / clinical code, investigation codes.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: d1/d2a/d2b, h4, h6.Mdr section codes updated/corrected: a.The revision reported was likely the result of failure of the patient¿s rotator cuff and four years of reported pain.The glenoid component was returned fractured with the center cage missing.The cause and sequence of events cannot be determined because no sequential, pre-revision radiographs were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|