MAUDE Adverse Event Report: VOLCANO CORPORATION VISIONS PV .018 DIGITAL IVUS CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VOLCANO CORPORATION VISIONS PV .018 DIGITAL IVUS CATHETER; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number 86700
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This case was reviewed and investigated according to the manufacturer¿s policy.Blocks a2-a5: information not provided.Block b6: information not provided.Block c: not applicable for this device.Blocks d6 & d7: not applicable for this device.Blocks h3 & h6: the visions pv.018 catheter was not returned for evaluation, thus no returned product investigation was performed.Blocks h7 & h9: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event or Problem Description
It was reported that a visions pv.018 catheter was used in a diagnostic peripheral procedure in a calcified distal at.While passing through the lesion, the distal tip separated but remained intact to the non-philips guidewire.All parts were removed successfully.The procedure was completed with another catheter.No patient injury reported.This product problem is being submitted because the visions distal tip separated.There is a potential for harm if the malfunction were to recur.
 
Additional Manufacturer Narrative
Blocks d9 & h3: the visions pv.018 catheter was returned for evaluation.Block h3: the visions pv.018 catheter was returned in two pieces.The distal tip separated, measuring approx.9 mm (0.9 cm).The proximal portion of the distal tip that was intact to the catheter, measured approx.3 mm (0.3 cm).The returned device measured 135.7 cm from the luer to the location of the separation.The expected overall catheter is 135-139 cm in length.Block h6: the probable cause of the tip separation is damage during use/handling.Device manipulation, impact and applied pressure associated with use and handling can further affect the integrity of the device.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Additional Manufacturer Narrative
Block d4: unique identifier (udi#) is corrected from (b)(4) to (b)(4).Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONS PV .018 DIGITAL IVUS CATHETER
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
VOLCANO CORPORATION
3721 valley centre drive #500
san diego CA 92130
Manufacturer (Section G)
VOLCARICA S.R.L.
coyol free zone &business park
b37
alajuela
CS  
Manufacturer Contact
ayse kharodawala
3721 valley centre drive #500
san diego, CA 92130
858720-4063
MDR Report Key19159431
Report Number3008363989-2024-00024
Device Sequence Number12721297
Product Code OBJ
UDI-Device Identifier00184360000037
UDI-Public(01)00184360000037(11)231102(17)251102(10)0303038976
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K150442
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/02/2025
Device Model Number86700
Device Catalogue Number86700
Device Lot Number0303038976
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
09/05/2024
Initial Report FDA Received Date04/22/2024
Supplement Report FDA Received Date05/03/2024
09/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ABBOTT: 0.014" COMMAND GUIDE WIRE.; PHILIPS: INTRASIGHT SYSTEM.; TERUMO: 5F SLENDER INTRODUCER SHEATH.
Patient SexUnknown
-
-