• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. NEXGEN 18MM DIAMETER 100MM LENGTH STRAIGHT STEM EXTENSION; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. NEXGEN 18MM DIAMETER 100MM LENGTH STRAIGHT STEM EXTENSION; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 03/27/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10 medical devices: nexgen 14mm height size c,d with locking screw yellow articular surface, catalog#: 00599403014, lot#: 62938369.Nexgen left size d cemented option femoral component, catalog#: 00599401491, lot#: 63746608.Nexgen stemmed tibial component precoat size 3, catalog#: 00598003701, lot#: 63544274.Nexgen 14mm diameter 100mm length offset stem extension, catalog#: 00598802014, lot#: 63632065.Trabecular metal cone full, catalog#: 00545004801, lot#: 63525256.Prc agmt block dist sz d 5mm, catalog#: 00599003410, lot#: 62963054.Prc agmt block dist sz d 5mm, catalog#: 00599003410, lot#: 62999435 competitor bone cement, catalog#: unk, lot#: unk.G2 foreign source: new zealand.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that patient underwent a left knee revision approximately six years post-implantation due to femoral pain.No additional patient consequences were reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Proposed component code: mechanical (g04) - stem.Reported event was unable to be confirmed due to limited information received from the customer.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: possible radiolucency at the tibial metal-cement interface and early loosening cannot be excluded.The bone has advanced osteoarthritis.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN 18MM DIAMETER 100MM LENGTH STRAIGHT STEM EXTENSION
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19159842
Report Number0002648920-2024-00115
Device Sequence Number15287600
Product Code JWH
UDI-Device Identifier00889024221024
UDI-Public(01)00889024221024(17)260228(10)63282086
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K933785
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberN/A
Device Catalogue Number00598801018
Device Lot Number63282086
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2024
Supplement Date Received by Manufacturer07/26/2024
Initial Report FDA Received Date04/22/2024
Supplement Report FDA Received Date08/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-