| Model Number |
75U |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection that the universal stopcock had multiple pieces of foreign material on the white plastic of the handle assembly and inside the stopcock body.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the foreign material was found to be three pieces of debris with two white colored and one yellow-brown colored.Further inspection also found pieces of white plastic and rust in the device.However, the root cause could not be identified.Olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
Although the device history record was unable to be reviewed for this device, the model number# was verified to be 75u and the lot# was bc as shown on the physical device.The root cause of the event was determined to be because of the presence of shavings inside the returned device which was a result of rework integrated into the assembly process.Olympus will continue to monitor field performance for this device.
|
| |
|