• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER-ENDOSCOPY LAKELAND 5300; LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER-ENDOSCOPY LAKELAND 5300; LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED Back to Search Results
Model Number B5LT
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the clear tip at the end of the complaint device broke off into the patient and, although the team did their did diligence in the or to look for it, it was not identified via ultrasound until after the patient left the or.It was additionally reported that the patient was brought back into the or a few days later to remove the piece during a second surgery.There was no other patient injury or medical intervention reported and the complainant is not aware of the extended procedure duration.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
The device was not returned to stryker sustainability solutions for evaluation.As the complaint device was not returned, a conclusive root cause could not be determined and the reported event could not be confirmed.No device information was reported and the customer did not report lot # or serial # information.Therefore, the device history record (dhr) was unable to be verified.The reported event could be attributed to: excessive force applied contact with hard object or other improper handling insufficient structural integrity shipping/handling damage or extreme conditions the instructions for use (ifu) state: - the reprocessed endoscopic trocars and sleeves have applications in abdominal, thoracic, and gynecologic minimally invasive surgical procedures to establish a path of entry for endoscopic instruments.The bladeless trocars may be used with or without visualization for primary and secondary insertions.Warnings - these instruments are only intended for use by individuals with adequate training and familiarity with minimally invasive techniques.For further information about techniques, complications and hazards, consult the medical literature.- prior to use, read and follow the instructions of this insert as well as those of the instruments and accessories to be used during the procedure.Damage to the instrument can lead to patient injuries.Always inspect instrument carefully for overall integrity before use.Directions for use - inspect the instrument and package before opening.The contents of the package are sterile if the packaging has not been compromised.If the package is damaged or if it was opened and the instrument was not used, return the instrument and packaging to stryker sustainability solutions for resterilization by ethylene oxide (eto) gas.- inspect the instruments for any damage.Do not use the instrument if any damage is noted.Return the instrument and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.- the trocar can be assembled by inserting the obturator into the cannula sleeve until they lock securely together.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED
Manufacturer (Section D)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key19160764
Report Number0001056128-2024-00040
Device Sequence Number7937856
Product Code NLM
UDI-Device Identifier00885825014193
UDI-Public00885825014193
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K201511
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberB5LT
Device Catalogue NumberB5LT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/27/2024
Initial Report FDA Received Date04/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-