STRYKER ORTHOPAEDICS-MAHWAH 32MM +4 LFIT V40 HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6260-9-232
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Pain (1994); Metal Related Pathology (4530)
Date of Event 03/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding abnormal ion level involving a metal head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinical review - no medical records were received for review with a clinical consultant.Product history review: review of the product history records indicate devices were manufactured and accepted into final stock with no reported relevant discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.H3 other text: device not returned.
 
Event or Problem Description
It was reported by claimant's counsel that the patient underwent right tha on (b)(6) 2016 and was implanted with an lfit v40 femoral head and an accolade ii hip stem.Subsequently she presented with increasing pain and workup showed elevated cobalt and chromium levels.She was revised on (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32MM +4 LFIT V40 HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
juana hiciano
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19162861
Report Number0002249697-2024-00626
Device Sequence Number7873558
Product Code JDI
UDI-Device Identifier07613327013085
UDI-Public07613327013085
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Catalogue Number6260-9-232
Device Lot Number53203702
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/28/2024
Initial Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-