• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VIANT MEDICAL, LLC EZ CLEAN REAMER HANDLE; STRAIGHT REAMER HANDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VIANT MEDICAL, LLC EZ CLEAN REAMER HANDLE; STRAIGHT REAMER HANDLE Back to Search Results
Model Number 00-1206-090-10
Medical Device Problem Codes Unstable (1667); Material Deformation (2976); Material Twisted/Bent (2981); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complaint sample was not returned to viant for evaluation.Thus, the reported event is non-verifiable.The current ifu sent with this device today, man-004006 rev.A, states the following; · end of life is determined by wear and damage due to intended use, · visually inspect for damage and wear.If the instrument is damaged and worn it is considered at the end of its life and should be discarded, · where instruments form part of a larger assembly, check assembly with mating components, · viant devices should only be used by qualified personnel fully trained in the use of the surgical instruments and the relevant surgical procedures, · do not modify viant instruments in any way and handle with care at all times.Surface scratches can increase wear and the risk of corrosion, · manual surgical instruments have a limited life-span which is determined by wear or damage due to repeated intended use.When a surgical instrument reaches the end of its functional life, clean the instrument of any and all biomaterial/biohazards and safely discard the instrument in accordance with applicable laws and regulations.The device history record (dhr) review was not performed as the lot number is unknown.It is unknown how many surgical procedures (cycles) this device has experienced throughout its life in the field as the lot number is unknown to determine the date of manufacture.The viant risk management files were reviewed to ensure the reported failure mode (or similar) is captured and assessed.The review revealed there are similar failure modes identified and mitigated to the lowest possible risk region.In conclusion, the complaint sample was not received by viant for evaluation and the reported event is non-verifiable.If the complaint sample is received by viant, it will be evaluated, the complaint record will be updated accordingly, and a supplemental medwatch 3500a emdr will be submitted as well.No further investigation with regard to this complaint is required at this time.Viant will continue to monitor for trends.Complaint information provided by distributor, zimmer biomet.
 
Event or Problem Description
It was reported during an unknown procedure of a 62 year old male patient that the reamer shaft wobbled and must have a slight bend in it.The case was completed with a second device.No consequences or impact to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ CLEAN REAMER HANDLE
Common Device Name
STRAIGHT REAMER HANDLE
Manufacturer (Section D)
VIANT MEDICAL, LLC
4545 kroemer road
fort wayne IN 46818
Manufacturer (Section G)
VIANT MEDICAL, LLC
4545 kroemer road
fort wayne IN 46818
Manufacturer Contact
tony singh
4545 kroemer road
fort wayne, IN 46818
MDR Report Key19163011
Report Number3004976965-2024-00008
Device Sequence Number7435656
Product Code LXH
UDI-Device Identifier00840096401394
UDI-Public00840096401394
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00-1206-090-10
Device Catalogue NumberT2756
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2024
Initial Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
-
-