| Model Number |
865040 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a speaker malfunction inop error message.It is unknown if there was still sound coming from the device.
|
| |
|
Additional Manufacturer Narrative
|
|
Reporting institution phone # (b)(6).Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips field service engineer (fse) went onsite to evaluate the device in question.Functional testing was performed and the results found a speaker inoperative message present and a speaker to be completely out of sound.Based on the information available and the testing conducted, the cause of the reported problem was confirmed to be a speaker issue.After the speak assembly was replaced, the unit returned to specification.No further investigation or action is warranted at this time.
|
| |
|