PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MP2 Back to Search Results
Model Number 865040
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported a speaker malfunction inop error message.It is unknown if there was still sound coming from the device.
 
Additional Manufacturer Narrative
Reporting institution phone # (b)(6).Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Additional Manufacturer Narrative
A philips field service engineer (fse) went onsite to evaluate the device in question.Functional testing was performed and the results found a speaker inoperative message present and a speaker to be completely out of sound.Based on the information available and the testing conducted, the cause of the reported problem was confirmed to be a speaker issue.After the speak assembly was replaced, the unit returned to specification.No further investigation or action is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MP2
Common Device Name
INTELLIVUE MP2
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
70314632031
MDR Report Key19163496
Report Number9610816-2024-00218
Device Sequence Number20658940
Product Code DSI
UDI-Device Identifier00884838000278
UDI-Public00884838000278
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K150310
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 06/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number865040
Device Catalogue Number865040
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/23/2024
Supplement Report FDA Received Date06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/14/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-