| Model Number |
MI1200 SYNCHRONY |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Codes |
Head Injury (1879); Failure of Implant (1924)
|
| Date of Event |
01/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a trauma.Revision surgery is considered.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a trauma.The user noticed a sudden change in hearing with the device.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: based on the received information, a damage to the active electrode due to the reported mechanical impact appears very likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: based on the received information, a damage to the active electrode due to the reported mechanical impact appears very likely.However, to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected after a trauma.The user noticed a sudden change in hearing with the device.The user was reimplanted.
|
| |
|