MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FEM TRIAL SZ 8 LT; FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FEM TRIAL SZ 8 LT; FEMORAL TRIALS Back to Search Results
Catalog Number 254500718
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
- procedure: total knee replacement it was reported that the trial femur broke during trialing - when did the even occur (pre-op/intra-op/post-op)? intra op - was there any patient consequence? no - what action was taken to resolve the issue? replace the trial femur - how long was the procedure prolonged as a result of the difficulties encountered with the device (minutes)? 10 minutes.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : procedure: total knee replacement; description of issue: the trial femur broke during trialing.The product was returned to depuy synthes for evaluation.Visual analysis of the returned device found that the attune cr fem trial sz 8 lt was broken at one of the condyles.The broken fragment was returned for evaluation.The fracture surface is consistent with overload due to a combination of heavy impaction and the trial fitting too tight over the posterior/anterior aspect of the resected femur bone forcing the curved features of the femoral trial to open.The combination of heavy impaction and possible discrepancies in the resection depth between the femur and the trial suggest unintended user error.Furthermore, per the attune surgical technique (dsus/jrc/0316/1437 rev.K) it is cautioned when extracting the trial, rocking the trial medio-laterally may cause condylar fracture.Such rocking should be avoided.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune cr fem trial sz 8 lt would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: corrected: reportability changed from serious injury to malfunction.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FEM TRIAL SZ 8 LT
Common Device Name
FEMORAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19163866
Report Number1818910-2024-08815
Device Sequence Number7817448
Product Code HWT
UDI-Device Identifier10603295133902
UDI-Public10603295133902
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500718
Device Lot NumberMVMGML730
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2024
Supplement Date Received by Manufacturer06/06/2024
06/18/2024
02/16/2026
Initial Report FDA Received Date04/23/2024
Supplement Report FDA Received Date06/10/2024
06/19/2024
02/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-