| Model Number |
GLENOID TARGETER SHAFT |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 3/28/2024, it was reported by a sales representative via sems-06526980 that an ar-5400-01 glenoid targeter shaft "b" slot was not locking down with a leg placed in it when setting up with the scrub tech.They tightened the capture nut and the leg, but would still not lock in.They tried using many legs of different sizes to see if this was the issue, but it persisted.The case was completed successfully, and no further information was provided.This was discovered when setting up with the scrub tech for a reverse total shoulder arthroplasty procedure on (b)(6) 2024, with no reported patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 complaint allegation is not confirmed.One unpackaged ar-5400-01 serial/batch number 022139 was received for investigation.However, the mating parts were not received.Visual evaluation noted nicks and the edges of the slots for the attachments.Functional testing found that the device does slide in and out of the matting part without issues.And the legs locked into place.No problem found.
|
| |
|