| Model Number |
ENSITE-SYSTEM-201 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a supraventricular tachycardia procedure, an unexpected shutdown occurred which delayed the procedure.Using navx mode v3.0.1, when mapping, creating and editing model, the system became slow and the dws shut down without any alerts being displayed.After shutting down, the red light on the dws's power button lit up three times and the beep sounded, which occurred repeatedly.The power outlet location was changed, and it was rebooted.Since it was rebooted successfully, the procedure continued, but under the same circumstances, this time it shut down without slowing down.The dws rebooted by disconnecting everything other than the power supply and restarting it, and the procedure continued.When the procedure resumed and rf delivery was performed, it shut down again.It was rebooted a third time the same way it was the second time.However, it did not reboot so fluoroscopic images and the recording system were used to check the location and effect of the rf delivery.The procedure was completed without replacing the ensite x with no consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One ensite x display workstation (dws) z4 was received for evaluation.Visual inspection revealed the front panel ports, and labels were free of physical damage.Internal visual inspection showed that all cables were secure and without pinch.The dws device accessories were connected along with power.Power was applied and the dws displayed an error message ¿928 fatal pcie error¿ on slot0¿.This is a power-on-self-test (post) symptom identified against the serial card installed on slot 0 of the motherboard.It was noted that 3 audible error beeps also relate to a hardware point, per the hewlett packard technical guide.Both symptoms are related and identify the serial card to be the cause.The device was left in the error state which the display workstation auto-powered down, duplicating the reported symptom.When the device came back up, the slot 0 error did repeat.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the investigation, and information provided to abbott, the reported event was able to be confirmed, and the root cause was attributed to the serial card installed in slot 0 of the motherboard.
|
| |
|
Search Alerts/Recalls
|
|
|