MAUDE Adverse Event Report: ARTHREX, INC. DYNANITE¿ STAPLE DRL BT, CAL, 1.6, REUSE; DRILL BIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. DYNANITE¿ STAPLE DRL BT, CAL, 1.6, REUSE; DRILL BIT Back to Search Results
Model Number DYNANITE¿ STAPLE DRL BT, CAL, 1.6, REUSE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint confirmed.Visual inspection identified that the drill bit shaft has broken off and tip is stuck inside the returned drill guide.A likely cause of the event is prying/leveraging the mating drill bit during use.
 
Event or Problem Description
It was reported that during a case, when drilling with the ar-8717d-01-ru, the drill locked in the ar-8717g-01 parallel guide.There was no adverse effect to the patient reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNANITE¿ STAPLE DRL BT, CAL, 1.6, REUSE
Common Device Name
DRILL BIT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19165145
Report Number1220246-2024-02253
Device Sequence Number7938006
Product Code HTW
UDI-Device Identifier00888867326163
UDI-Public00888867326163
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDYNANITE¿ STAPLE DRL BT, CAL, 1.6, REUSE
Device Catalogue NumberAR-8717D-01-RU
Device Lot Number031938
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2021
Initial Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/01/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-