| Model Number |
N/A |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign ¿ canada.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, the liner could not be placed in the cup.A new liner was used to complete the procedure.There was no known impact or consequences to the patient.It was reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination of the returned product/provided pictures identified gouges and indentations to the rim, inner spherical surface, and elevated anti rotation scallops.Cut outs on the outer spherical surface show indentations and deformation from attempted implanting.No other damage was noted.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.This complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|