EXACTECH, INC. NV GXL LNR, +5LAT, 32MM G1-48/50MM CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 136-32-51
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 02/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitant devices: 2135060 186-01-50 - integrip cc, cluster 50mm, g1; 2538871 101-45-40 - 4.5mm drill bit 40mm; 2563052 122-65-25 - 6.5mm acetabular bone screw 25mm.H7: recall number z-1729-2022.H3: the revision reported was likely the result of prosthesis wear of the tibial insert.The cause of prosthesis wear is generally a combination of risk factors including use error, implant positioning, implant size selection, and patient factors.However, this cannot be conclusively determined with the information provided.
 
Event or Problem Description
As reported via legal documentation, a patient had a left total hip replacement procedure on (b)(6) 2013 and then was revised on (b)(6) 2023.The patient required revision surgery for issues including but not limited to polyethylene prosthesis wear, component loosening, device failure.The mostly likely cause for the revision reported due to prosthesis wear is a combination of risk factors including use error, implant positioning, implant size selection, and patient factors.However, this cannot be conclusively determined with the information provided.
 
Additional Manufacturer Narrative
B1: corrected.B2: corrected.H6: corrected, updated the following: health effect - clinical code, type of investigation.
 
Additional Manufacturer Narrative
H3 corrected: the most likely cause for the reported revision due to prosthesis wear is a combination of the risk factors specified in hhe.However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, +5LAT, 32MM G1-48/50MM CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
matt collins
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key19167667
Report Number1038671-2024-00955
Device Sequence Number17987497
Product Code JDI
UDI-Device Identifier10885862024275
UDI-Public10885862024275
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K150890
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2015
Device Catalogue Number136-32-51
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date04/23/2024
Supplement Report FDA Received Date09/04/2024
09/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/04/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-