Catalog Number 7210980
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference number: (b)(4).
 
Event or Problem Description
It was reported that during a knee arthroscopy, the shaver got clogged, tech went to unclog it and noticed the tip on the inside of the full radius blade had broken off inside the shaver sheet.No pieces were left in the patient, they were not in the patient freely.Another blade was opened and the same thing happened.The procedure was completed without surgical delay using a smith and nephew back up device.No further complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.A visual inspection revealed that they are not in their original packaging.The inner shaft of both devices is fractured near the distal end of the heat shrink sleeve.The distal blade portion of the inner shafts is stuck in the outer shaft of the devices.There is biological debris on the returned items.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include an inadvertent impact event inconsistent with normal use or an application of unintended inappropriate or excessive force to the device.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLADE 3.5 CURVED CONCAVE FULL RADIUS
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19168676
Report Number1219602-2024-00865
Device Sequence Number20678123
Product Code HAB
UDI-Device Identifier03596010595775
UDI-Public03596010595775
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number7210980
Device Lot Number51169563
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2024
Supplement Date Received by Manufacturer07/05/2024
Initial Report FDA Received Date04/23/2024
Supplement Report FDA Received Date06/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-