ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAP-0700
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Date 04/10/2024
Event Type  malfunction  
Event Description
It was reported "pump had interference on ap waveform".No patient harm or injury reported.Additionally, it was reported that the pump was not switched out.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
Qn# (b)(4).The reported complaint of "interference on ap waveform" could not be confirmed upon investigation of the returned sample.The customer returned an ac3 i/o board, a front-end board, and a cpm board for investigation.The returned parts passed visual and functional test specifications during the complaint investigation.Based on a review of the device history record (dhr), the product met specification upon release.No problem was found with the returned device.No further action required at this time.Teleflex will continue to monitor and trend on complaints of this nature.
 
Event Description
It was reported "pump had interference on ap waveform".No patient harm or injury reported.Additionally, it was reported that the pump was not switched out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville 27560
MDR Report Key19168774
MDR Text Key340964943
Report Number3010532612-2024-00343
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902084966
UDI-Public30801902084966
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/10/2024
Initial Date FDA Received04/23/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/30/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
NONE REPORTED
-
-