• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ECHELON CONTOUR W/ BLUE RELOAD; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ECHELON CONTOUR W/ BLUE RELOAD; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number GCS40B
Medical Device Problem Codes Difficult to Open or Close (2921); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a colostomy procedure the contour retaining pin wasn't going forward where the device couldn't close/be used.They got another device to complete the colostomy without patient harm.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 4/23/2024.D4: batch #452a94.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the gcs40b device was received with no apparent damage and with a reload loaded in the device.The reload was received fully loaded with staples, with the washer uncut, with the drivers and with the knife recess below the reload deck.The reload was not properly loaded in the device, as the retaining pin was not connected to the coupler and pushrod.These facts indicate that the reload was removed and reinserted incorrectly and/or incompletely after the retainer was removed.The returned reload was pulled out of the device and placed back on; the device opened and closed without any difficulties.The device was tested for functionality with a returned reload and the device fired, forming all staples and cutting as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria.It should be noted that if the reload is removed from the device, whether the reload is spent or not, and if the staple retainer has already been removed from the reload, the reload cannot be reloaded into the device.Please reference the instructions for use for the complete guide loading and reloading the instrument.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 452a94, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device gcs40b lot number v95z68 and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON CONTOUR W/ BLUE RELOAD
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
PR  
3035526892
MDR Report Key19169079
Report Number3005075853-2024-03180
Device Sequence Number15441044
Product Code GDW
UDI-Device Identifier10705036027515
UDI-Public10705036027515
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K200420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Catalogue NumberGCS40B
Device Lot Number452A94
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/22/2024
Initial Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-