• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE CAGE GLENOID M, POST AUG, RIGHT; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE CAGE GLENOID M, POST AUG, RIGHT; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number EQUINOXE CAGE GLENOID M, POST AUG, RIGHT
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Osteolysis (2377)
Date of Event 03/07/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported, approximately five years post initial right tsa, the female patient had a failure of anatomic shoulder replacement.Patient had a revision of legacy caged glenoid, the central cage appeared to have disassociated from the poly.There were no fragments, parts or pieces that fell into the patient wound site.There was no surgical delay/prolongation.Patient was last known to be in stable condition following the event.Images received.Sales rep was unable to obtain x-rays.The device is not available for evaluation due to lost after case.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6 mdr section codes updated/corrected: b, c, d, e the revision reported was likely the result of prosthesis fracture of the glenoid.The cause of the fracture may be due to incomplete seating of the implant and/or off-axis drilling of the peripheral peg holes during implantation resulting in a rocking horse effect.However, the series of events could not be confirmed as the device was not returned for evaluation and no radiographs were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE CAGE GLENOID M, POST AUG, RIGHT
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
geoff gannon
MDR Report Key19169322
Report Number1038671-2024-00957
Device Sequence Number7304134
Product Code KWS
UDI-Device Identifier10885862201010
UDI-Public10885862201010
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K113309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/30/2023
Device Model NumberEQUINOXE CAGE GLENOID M, POST AUG, RIGHT
Device Catalogue Number314-13-33
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/23/2024
Supplement Report FDA Received Date09/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EQUINOX SQUARE TORQUE DEFINE SCREW DRIVE KIT.; EQUINOXE CAGE GLENOID M, POST AUG, RIGHT.; EQUINOXE REPLICATOR PLATE 1.5MM O/S.; EQUINOXE, HUMERAL HEAD SHORT, 44MM (ALPHA).; EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT.; GPS IMPLANT KIT V2.; GPS PER CASE RENTAL FEE56.; SHOULDR GPS HEX PINS KIT.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-