| Catalog Number |
72201514 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee arthroscopy, there were metal shavings or fragments noted while using the incisor plus elite.All fragments removed by suction.The procedure was successfully completed without surgical delay using the same faulty device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.H11: corrected information in h6 (health effect - impact code).
|
| |
|