Model Number BEA25-110/I20-30
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Codes Aneurysm (1708); Clinical Endoleak (4843)
Date of Event 04/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was treated for an abdominal aortic aneurysm (aaa) on (b)(6) 2023, with the implantation of an afx2 bifurcated stent graft, an afx vela infrarenal, and an afx vela suprarenal.It was reported on (b)(6) 2024, that the patient underwent a re-intervention for an endoleak type ib.The physician implanted an additional afx2 bifurcated stent graft and an ovation ix extender.The final patient status was reported as doing well with no presence of an endoleak.
 
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.H3 other text : devices remain implanted.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix's practice to make at least three good-faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device was properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix confirms the afx2 type ib endoleak of the right common iliac artery and additional endovascular procedure complaints.This is consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest an aneurysm enlargement of 5 mm.The aneurysm enlargement was discovered during the review of the computed tomography (ct) scan dated on (b)(6) 2023.Procedure-related harms, device, user, procedure, or anatomy-relatedness of this complaint could not be determined with the medical records available for review.The final patient status was reported as doing well.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g3: awareness date ¿ updated h6: investigation finding codes - remove code 3233 h6: investigation conclusion codes - remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key19169591
Report Number3011063223-2024-00060
Device Sequence Number12599965
Product Code MIH
UDI-Device Identifier00818009014818
UDI-Public(01)00818009014818(17)240908
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/08/2024
Device Model NumberBEA25-110/I20-30
Device Lot Number2505375-005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/01/2024
Supplement Date Received by Manufacturer05/22/2024
Initial Report FDA Received Date04/23/2024
Supplement Report FDA Received Date06/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA INFRARENAL 2361261007; AFX VELA SUPRARENAL 2578207020
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexMale
-
-