| Catalog Number |
392533 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
03/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd iag bc pro incomplete needle retraction.The following information was provided by the initial reporter: safety mechanism did not work as intended.Needle did not retract all the way into the housing leaving the contaminated sharp exposed.
|
| |
|
Event or Problem Description
|
|
Additional information from customer: there was 2 needlesticks to nurses.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: the complaint that the needle failed to fully retract was confirmed; however, the root cause could not be determined from the photographs that were provided for investigation.One photo showed just a short section of the needle extending from the grip, which indicated that the safety mechanism was activated and the needle was partially retracted.This type of damage can occur if the barrel is damaged, which prevents full retraction of the needle hub and needle.The barrel is being held in the photo, which obscured a portion of the needle hub and barrel.The root cause of the reported issue could not be determined from the photo.A device history record review showed no non-conformances associated with this issue during the production of the reported production batches.Complaints received for this device and reported condition will continue to be tracked and trended.
|
| |
|