VYAIRE MEDICAL M2M ASSY, PKG, REVEL, GRN, DISS, ENGL; VENTILATOR, CONTINUOUS, FACILITY USE
|
Back to Search Results |
|
| Model Number |
REVEL, GRN, DISS, ENGL |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to vyaire medical that upon going to start up the vent it began beeping and vent inop light is seeing flashing red along with the control lock, manual breath, and maneuvers buttons all being lit up green.The vent cycled on twice with the led screen lighting up for a second but then shutting off (going blank) immediately after both times.No patient involvement, issue found during start-up.
|
| |
|
Additional Manufacturer Narrative
|
|
Vyaire medical file identification:(b)(4).H3: other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: finding/root-cause: reported complaint was verified and duplicated.This was isolated to capacitors p/n:13804-001 and has been addressed by capa-000000497.
|
| |
|