• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDEX® PEDIATRIC ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDEX® PEDIATRIC ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 165810
Medical Device Problem Code Inaccurate Flow Rate (1249)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the urine did not drain from the foley catheter.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The reported event is unconfirmed.No root cause could be found because the reported event was unconfirmed.Photo: received four (4) photo samples.Three (3) photo samples showcase an overview of an all-silicone foley catheter.Fourth photo sample showcases a piece of paper with native writing.Visual: received one (1) used all-silicone foley catheter without original packaging.Visual inspection noted no obvious observations.The catheter balloon was inflated with 5ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and the catheter rested with no leaks noted.The balloon passively deflated in under two (2) minutes with no cuffing or leaks noted, returning 5ml of solution.This is within specification which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.As the reported event is unconfirmed a dhr review is not required.However, a dhr was completed before unconfirming the event.A labeling review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the urine did not drain from the foley catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® PEDIATRIC ALL-SILICONE FOLEY CATHETER
Common Device Name
SILICONE FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key19169876
Report Number1018233-2024-02244
Device Sequence Number15299338
Product Code KOD
UDI-Device Identifier00801741029721
UDI-Public(01)00801741029721
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K821745
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number165810
Device Lot NumberNGHS3190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2024
Supplement Date Received by Manufacturer06/13/2024
Initial Report FDA Received Date04/24/2024
Supplement Report FDA Received Date06/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-