| Catalog Number |
8065149527 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A nurse reported that during surgery an ophthalmic surgical soft tip in the cannula came off while being injected the product inside the patient's eye (unknown) and it was retrieved with a forceps.The surgery was completed on the same day.The type of procedure and patient impact was not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A sample was not received at the manufacturing site for evaluation for the report of plastic tip in the cannula came off; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the product was processed and released according to the product¿s acceptable criteria, therefore, the root cause for customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.All assemblies are 100% inspected for the soft tip during the manufacturing process to ensure that the product meets release acceptance criteria.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information has been received and it was reported that the procedure was completed on the same day after replacing with new cannula.No patient impact was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information was provided in section d4.The manufacturer internal reference number is: (b)(4).
|
| |
|