Catalog Number 136-36-53
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330)
Date of Event 05/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitants: 101-05-20 - 3.2mm drill bit 20mm 1pk 5542263 170-36-93 - biolox delta femoral head 36mm od, -3.5mm 5593255 180-65-20 - alteon 6.5mm screw, 20mm 3728749 180-65-20 - alteon 6.5mm screw, 20mm 5196613 186-01-56 - integrip cc, cluster 56mm,g3 5512778 190-30-06 - alt ha s clr std sz 6 5207651.These devices are used for treatments, not diagnosis.There is no other information available.
 
Event or Problem Description
It was reported via legal documentation that a patient had a left total hip arthroplasty on (b)(6) 2018, and then experienced revision surgical procedure on (b)(6) 2022 approximately 3 years and 7 months after initial implant.No images were provided.There is no other information available.
 
Additional Manufacturer Narrative
H3: based on the available information, the patient involved meets the following risk criteria for early prosthesis wear and/or osteolysis as specified in the hhe: implanted with a lateralized liner.The most likely cause for the reported revision due to prosthesis wear is a combination of the risk factors specified in hhe.However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b5, e4: removed erroneous entry, g2: consumer, h6: clinical code.Based on the available information, the patient involved meets the following risk criteria for early prosthesis wear and/or osteolysis as specified: implanted with a lateralized liner.The most likely cause for the reported revision due to prosthesis wear is a combination of the risk factors.However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that the patient underwent an initial left total hip arthroplasty.Subsequently, the patient experienced pain, discomfort and prosthesis wear.As a result, the patient underwent a left hip revision procedure approximately 3 years and 7 months after initial implantation.No further patient impact reported.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, +5LAT, 36MM G3-56/58MM CUPS
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROU
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
matt collins
MDR Report Key19170232
Report Number1038671-2024-00959
Device Sequence Number7856239
Product Code LZO
UDI-Device Identifier10885862024336
UDI-Public10885862024336
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/28/2022
Device Catalogue Number136-36-53
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2023
Supplement Date Received by Manufacturer08/14/2024
02/03/2026
Initial Report FDA Received Date04/24/2024
Supplement Report FDA Received Date08/20/2024
02/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/30/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1729-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-