EXACTECH, INC. NV GXL LINR, NTRL, 36MM ID, GROUP 4 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROU
|
Back to Search Results |
|
| Catalog Number |
130-36-54 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Distress (2329); Osteolysis (2377); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/11/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10.D10.Concomitants: 4118946 - 164-01-12 - element-stem, collarless w/ha, std offset, sz 12 4487770 - 170-36-07 - biolox delta femoral head 36mm od, +7mm 4414694 - 186-03-60 - integrip mh cup sz 60mm.These devices are used for treatments, not diagnosis.There is no other information available.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that a patient had a left total hip arthroplasty on (b)(6) 2016, and then experienced revision surgical procedure on (b)(6) 2023 approximately 6 years and 9 months after initial implant.No images were provided.There is no other information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information, the patient meets the following risk criteria for early prosthesis wear and/or osteolysis: implanted with a lateralized liner, combination of largest available femoral head and/or thinnest available acetabular liner was used, and implanted with a component having a shelf age of greater than 2 years.The most likely cause for the reported revision due to prosthesis wear and osteolysis may have been a combination of risk factors such as use error, implant positioning, implant size selection, and patient factors (fitness for surgery, biomechanics, activity level and local tissue oxidation potential).However, this cannot be confirmed as the devices were not returned for evaluation, and images and radiographs were not provided.H6: corrected health effect and investigation clinical codes.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: corrected health effect - clinical code.
|
| |
|
Search Alerts/Recalls
|
|
|