| Model Number |
866470 |
| Medical Device Problem Code |
No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the system had a technical alarm loudspeaker error.It is unknown if the device was in use at time of event, and there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips field service engineer (fse) went onsite to evaluate the device in question.The fse indicated during an evaluation of the system that the system logs displayed the speaker malfunction.The fse provided the customer with a replacement device speaker to repair a device.The philips field service engineer (fse) confirmed the customer's device issue customer with a replacement device speaker to repair the device.(b)(6).
|
| |
|